Operation:
The 4115 is easy to operate. The outside air calibration should be
adjusted to 100% via a supplied screwdriver. Initially if the unit
displays 99% (and is not adjusted to 100%) then the tolerance
will then be 1% of the final value.
For example, if the final value is
40 ppm, the difference is 1% of 40 ppm which is 0.4% ppm. So a variance in
the reading for the outside air results in only a very small change in
the final ppm read-out. The actual internal unit calibration is factory
set with certified gas and should never be adjusted.
To get a reading, attach the supplied probe to the hose and turn the
pump on. After use, turn pump off and return to case.
Optional bulkhead fittings are also available.
Maintenance:
The Aero2-Mat 4115 Series is relatively maintenance-free, but there
are some steps that should be taken to ensure continued reliability.
Both models use the same electode and electrolyte.
After 6 months of normal use, the electrolyte should be changed.
When the electrolyte is exhausted, the calibration will no longer
be possible and a reading of 150% will be displayed.
The outside air filter should be changed when any discoloration is noticed
to prevent possible damage from dust on the job site. These
parts are user replaceable or can be done in-house at Intercon.
>New manuals are available if the original is lost or destroyed.